Flame Photometry: Its Principle, Instrumentation & Applications

Flame photometry is a process where in emission of radiation by neutral atoms is measured.

The neutral atoms are obtained by introduction of sample into flame. Hence the name flame photometry.

Since radiation is emitted it is also called as flame emission spectroscopy.

Flame photometer working principle:

When a solution of metallic salt is sprayed as fine droplets into a flame. Due to heat of the flame, the droplets dry leaving a fine residue of salt. This fine residue converts into neutral atoms.

Due to the thermal energy of the flame, the atoms get excited and there after return to ground state. In this process of return to ground state, exited atoms emit radiation of specific wavelength. This wavelength of radiation emitted is specific for every element.

This specificity of wavelength of light emitted makes it a qualitative aspect. While the intensity of radiation, depends on the concentration of element. This makes it a quantitative aspect.
The process seems to be simple and applicable to all elements. But in practice only a few elements of Group IA and group IIA  (like Li, Na, k & Ca, Mg) are only analyzed.
The radiation emitted in the process is of specific wavelength. Like for Sodium (Na) 589nm yellow radiation, Potassium 767nm range radiation.

Flame photometer instrumentation:

1. Burner
2. Monochromators
3. Detectors
4. Recorder and display.

Burner: This is a part which produces excited atoms. Here the sample solution is sprayed into fuel and oxidant combination. A homogenous flame of stable intensity is produced.

Total Consumption burner

Total Consumption burner

There are different types of burners like Total consumption burner, Laminar flow and Mecker burner.

Fuel and oxidants: Fuel and oxidant are required to produce flame such that the sample converts to neutral atoms and get excited by heat energy.
The temperature of flame should be stable and also ideal. If the temperature is high, the elements in sample convert into ions instead of neutral atoms. If it is too low, atoms may not go to exited state. So a combination of fuel and oxidants is used such that there is desired temperature.
The following combination of fuel and oxidants are commonly used.

FuelOxidantTemperature of Flame
Propane +Air2100 Degree C
Propane +Oxygen2800 Degree C
Hydrogen +Air1900 Degree C
Hydrogen +Oxygen2800 Degree C
Acetylene +Air2200 Degree C
Acetylene +Oxygen3000 Degree C

Filters and monochromators are needed to isolate the light of specific wavelength from remaining light of the flame. For this simple filters are sufficient as we study only few elements like Ca, Na, K and Li. So a filter wheel with filter for each element is taken. When a particular element is analyzed, the particular filter is used so that it filters all other wavelengths.

Detector: Flame photometric detector is similar to that used in spectrophotometry. The emitted radiation is in the visible region i.e. 400nm to 700nm. Further the radiation is specific for each element so simple detectors are sufficient for the purpose like photo voltaic cells, photo tubes etc.

Recorders and display: These are he devices to read out the recording from detectors.

flame photometer

Flame photometer uses/ Applications
1. For qualitative analysis of samples by comparison of spectrum emission wavelengths with that of standards.
2. For quantitative analysis to determine the concentration of group IA and IIA elements. For example
a) Concentration of calcium in hard water.
b) Concentration of Sodium, potassium in Urine
c) Concentration of calcium and other elements in bio-glass and ceramic materials.

Flame photometry limitations:

Unlike other spectroscopy methods, flame photometry finds little use in research and analysis. This is due to

  1. Limited number of elements that can be analyzed.
  2. The sample requires to be introduced as solution into fine droplets. Many metallic salts, soil, plant and other compounds are insoluble in common solvents. Hence, they can’t be analyzed by this method.
  3. Since sample is volatilized, if small amount of sample is present, it is tough to analyze by this method. As some of it gets wasted by vaporization.
  4. Further during solubilisation with solvents, other impurities might mix up with sample and may lead to errors in the spectra observed.